// CASE EXAMPLES

Selected cases.

De-identified examples from retrospective analyses of pediatric spinal deformity and adult cervical spine surgery, chosen to illustrate how SureStim's view of the signal can differ from what the monitoring team sees. They are selected illustrations and are not representative of overall performance.

HOW TO READ THE CHARTS
Similarity. How closely the patient's current signals match their own baseline (dashed line). A line falling below it means the signal is drifting from normal for that patient.
Amplitude. Motor signal strength as a percent of baseline (dashed line), the measure most teams act on. Lower limb for pediatric cases, upper limb for adult cervical cases.
Team alert. When the monitoring team recorded the first change. The green marker shows when SureStim flagged elevated risk. White labels show the change from baseline. Cases tagged NO TEAM ALERT are ones where the monitoring team did not report a change.
CASE 1 · PEDIATRIC
16M congenital scoliosis s/p T2–S1 PSIF
34 min earlier
MOTOR WEAKNESS
CASE 2 · PEDIATRIC
16F congenital scoliosis s/p PSIF
28 min earlier
MOTOR WEAKNESS
CASE 3 · PEDIATRIC
16M AIS s/p T2–L4 PSIF
86 min earlier
NO DEFICIT
CASE 4 · PEDIATRIC
19M neuromuscular scoliosis s/p T4–sacrum PSIF
20 min earlier
NO DEFICIT
CASE 5 · ADULT CERVICAL
60M cervical myelopathy s/p C3–C5 ACDF
51 min earlier
POSTOP WEAKNESS
CASE 6 · ADULT CERVICAL
78M cervical myelopathy s/p C3–C7 laminectomy / laminoplasty
15 min earlier
POSTOP WEAKNESS
CASE 7 · ADULT CERVICAL
62F cervical myelopathy s/p C3–C7 ACDF
NO TEAM ALERT
POSTOP WEAKNESS
CASE 8 · ADULT CERVICAL
55F cervical myelopathy s/p C2–C5 laminectomy / laminoplasty
Similarity down, amplitude up
IMPROVED MOTOR STRENGTH
CASE 9 · ADULT CERVICAL
71F cervical myelopathy with myelomalacia s/p C4–C7 ACDF
Similarity down, amplitude up
IMPROVED MOTOR STRENGTH

Case 1

PEDIATRIC SPINAL DEFORMITYNEW POSTOP MOTOR WEAKNESS
16M congenital scoliosis s/p T2–S1 PSIF

By the team alert, similarity was 69% below this patient's own baseline and amplitude had fallen to 18% of baseline. The SureStim flag came 34 minutes earlier, while the signal still looked acceptable on amplitude alone.

34 min
SureStim flag before the team alert
−69%
Similarity vs own baseline at the team alert
18%
Lower-limb amplitude vs baseline at the team alert
Motor weakness
New weakness after surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% OWN BASELINE SURESTIM FLAG −34 MIN TEAM ALERT 0 −35% −69%
LOWER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 500% 100% BASELINE −93% −180 −150 −120 −90 −60 −30 0 +30 +60 MINUTES RELATIVE TO TEAM ALERT

Case 2

PEDIATRIC SPINAL DEFORMITYNEW POSTOP MOTOR WEAKNESS
16F congenital scoliosis s/p PSIF

By the team alert, similarity was 33% below this patient's own baseline and amplitude had fallen to 9% of baseline. The SureStim flag came 28 minutes earlier, while the signal still looked acceptable on amplitude alone.

28 min
SureStim flag before the team alert
−33%
Similarity vs own baseline at the team alert
9%
Lower-limb amplitude vs baseline at the team alert
Motor weakness
New weakness after surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% OWN BASELINE SURESTIM FLAG −28 MIN TEAM ALERT 0 −39% −33%
LOWER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 500% 100% BASELINE −95% −180 −150 −120 −90 −60 −30 0 +30 +60 MINUTES RELATIVE TO TEAM ALERT

Case 3

PEDIATRIC SPINAL DEFORMITYNO POSTOP DEFICIT
16M AIS s/p T2–L4 PSIF

SureStim flagged a sustained fall in similarity 86 minutes before the team recorded its first MEP change, while lower-limb amplitude was still near baseline. Within three minutes of that first recorded change, amplitude fell below half of baseline and bottomed out near 3% about 14 minutes in, around when the surgeon was alerted (+13 min), before recovering. Similarity was already about 50% below this patient's own baseline when the team first recorded a change.

86 min
SureStim flag before the team alert
−51%
Similarity vs own baseline at the team alert
27%
Lower-limb amplitude vs baseline at the team alert
No deficit
True signal change, no deficit after surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% OWN BASELINE SURESTIM FLAG −86 MIN TEAM ALERT 0 SURGEON ALERTED +13 −51%
LOWER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 500% 100% BASELINE −73% −97% −180 −150 −120 −90 −60 −30 0 +30 +60 MINUTES RELATIVE TO TEAM ALERT

Case 4

PEDIATRIC SPINAL DEFORMITYNO POSTOP DEFICITSIMILARITY DOWN · AMPLITUDE UP
19M neuromuscular scoliosis s/p T4–sacrum PSIF

At the team alert, lower-limb amplitude was still at 119% of baseline, while similarity had already fallen 37% below this patient's own baseline. The SureStim flag came 20 minutes before the team alert.

20 min
SureStim flag before the team alert
−37%
Similarity vs own baseline at the team alert
119%
Lower-limb amplitude vs baseline at the team alert
No deficit
True signal change, no deficit after surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% OWN BASELINE SURESTIM FLAG −20 MIN TEAM ALERT 0 −42% −37%
LOWER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 500% 100% BASELINE +19% −180 −150 −120 −90 −60 −30 0 +30 +60 MINUTES RELATIVE TO TEAM ALERT

Case 5

ADULT CERVICAL SPINENEW POSTOP MOTOR WEAKNESSSIMILARITY DOWN · AMPLITUDE UP
60M cervical myelopathy s/p C3–C5 ACDF

SureStim flagged a sustained fall in similarity 51 minutes before the team recorded a change. At the flag, similarity was 43% below this patient's own baseline while upper-limb amplitude was still at 159% of baseline, so the change was not visible on amplitude alone.

51 min
SureStim flag before the team alert
−43%
Similarity vs own baseline at the flag
159%
Upper-limb amplitude vs baseline at the flag
Motor weakness
New weakness immediately after surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% BASELINE SETS OWN BASELINE SURESTIM FLAG TEAM ALERT −43%
UPPER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 1000% 100% BASELINE +59% 0 60 120 180 240 300 MINUTES FROM FIRST RECORDED SIGNAL SET

Case 6

ADULT CERVICAL SPINENEW POSTOP MOTOR WEAKNESSSIMILARITY DOWN · AMPLITUDE UP
78M cervical myelopathy s/p C3–C7 laminectomy / laminoplasty

SureStim flagged a sustained fall in similarity 15 minutes before the team recorded a change. At the flag, similarity was 40% below this patient's own baseline while upper-limb amplitude was still at 163% of baseline, so the change was not visible on amplitude alone.

15 min
SureStim flag before the team alert
−40%
Similarity vs own baseline at the flag
163%
Upper-limb amplitude vs baseline at the flag
Motor weakness
New weakness by postop day 2
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% BASELINE SETS OWN BASELINE SURESTIM FLAG TEAM ALERT −40% −64%
UPPER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 1000% 100% BASELINE +63% +106% 0 60 120 180 240 MINUTES FROM FIRST RECORDED SIGNAL SET

Case 7

ADULT CERVICAL SPINENO TEAM ALERTNEW POSTOP MOTOR WEAKNESS
62F cervical myelopathy s/p C3–C7 ACDF

The monitoring team did not record an alert in this case. SureStim's similarity measure fell to 57% below this patient's own baseline about 219 minutes into monitoring. Upper-limb amplitude was at 29% of baseline at that point.

NO TEAM ALERT. The monitoring team did not report a signal change in this case, so there is no team alert marker on the charts.
No alert
Team recorded no signal change
−57%
Largest similarity drop vs own baseline
29%
Upper-limb amplitude vs baseline at that point
Motor weakness
New weakness by postop day 1
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% BASELINE SETS OWN BASELINE SURESTIM FLAG NO TEAM ALERT RECORDED −57%
UPPER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 1000% 100% BASELINE −71% 0 60 120 180 240 300 MINUTES FROM FIRST RECORDED SIGNAL SET

Case 8

ADULT CERVICAL SPINEIMPROVED MOTOR STRENGTHSIMILARITY DOWN · AMPLITUDE UP
55F cervical myelopathy s/p C2–C5 laminectomy / laminoplasty

Similarity to this patient's own baseline fell as much as 92% while upper-limb amplitude rose to 1065% of baseline, finishing the case at about 716%. Arm strength was better at the first postop exam than before surgery. A fall in similarity together with a rise in amplitude can mean the signals are stronger than at baseline but shaped differently from it, which after decompression is consistent with improved conduction and strength rather than injury. This is a hypothesis from a small number of cases, not a validated indicator.

−92%
Largest similarity drop vs own baseline
1065%
Upper-limb amplitude vs baseline at that point
716%
Upper-limb amplitude vs baseline at end of case
Improved motor strength
Arm strength better at the first postop exam than before surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% BASELINE SETS OWN BASELINE −92%
UPPER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 1000% 100% BASELINE +965% 0 60 120 180 240 MINUTES FROM FIRST RECORDED SIGNAL SET

Case 9

ADULT CERVICAL SPINEIMPROVED MOTOR STRENGTHSIMILARITY DOWN · AMPLITUDE UP
71F cervical myelopathy with myelomalacia s/p C4–C7 ACDF

Similarity to this patient's own baseline fell as much as 49% while upper-limb amplitude rose to 586% of baseline, finishing the case at about 630%. Arm strength was better at the first postop exam than before surgery. A fall in similarity together with a rise in amplitude can mean the signals are stronger than at baseline but shaped differently from it, which after decompression is consistent with improved conduction and strength rather than injury. This is a hypothesis from a small number of cases, not a validated indicator.

−49%
Largest similarity drop vs own baseline
586%
Upper-limb amplitude vs baseline at that point
630%
Upper-limb amplitude vs baseline at end of case
Improved motor strength
Arm strength better at the first postop exam than before surgery
SIMILARITY TO THE PATIENT'S OWN BASELINE
0% 25% 50% 75% 100% BASELINE SETS OWN BASELINE −49%
UPPER-LIMB MOTOR AMPLITUDE, % OF BASELINE (LOG SCALE)
10% 1000% 100% BASELINE +486% 0 60 120 180 240 MINUTES FROM FIRST RECORDED SIGNAL SET

Dates and identifying details have been removed. Pediatric cases were selected from the 14 patients with a true neurological change in a retrospective cohort, with times relative to the team alert. Adult cervical cases were selected from a retrospective cohort of 81 patients, with times measured from the first recorded signal set; the adult flag threshold was set on the same data. SureStim flag times come from a retrospective run of the model. Baseline similarity is measured on each patient's own early signals. These cases are not a measure of typical or prospective performance. SureStim is an investigational technology.

Want to see more of the data?

Partners and clinicians can request a full briefing.