De-identified examples from retrospective analyses of pediatric spinal deformity and adult cervical spine surgery, chosen to illustrate how SureStim's view of the signal can differ from what the monitoring team sees. They are selected illustrations and are not representative of overall performance.
By the team alert, similarity was 69% below this patient's own baseline and amplitude had fallen to 18% of baseline. The SureStim flag came 34 minutes earlier, while the signal still looked acceptable on amplitude alone.
By the team alert, similarity was 33% below this patient's own baseline and amplitude had fallen to 9% of baseline. The SureStim flag came 28 minutes earlier, while the signal still looked acceptable on amplitude alone.
SureStim flagged a sustained fall in similarity 86 minutes before the team recorded its first MEP change, while lower-limb amplitude was still near baseline. Within three minutes of that first recorded change, amplitude fell below half of baseline and bottomed out near 3% about 14 minutes in, around when the surgeon was alerted (+13 min), before recovering. Similarity was already about 50% below this patient's own baseline when the team first recorded a change.
At the team alert, lower-limb amplitude was still at 119% of baseline, while similarity had already fallen 37% below this patient's own baseline. The SureStim flag came 20 minutes before the team alert.
SureStim flagged a sustained fall in similarity 51 minutes before the team recorded a change. At the flag, similarity was 43% below this patient's own baseline while upper-limb amplitude was still at 159% of baseline, so the change was not visible on amplitude alone.
SureStim flagged a sustained fall in similarity 15 minutes before the team recorded a change. At the flag, similarity was 40% below this patient's own baseline while upper-limb amplitude was still at 163% of baseline, so the change was not visible on amplitude alone.
The monitoring team did not record an alert in this case. SureStim's similarity measure fell to 57% below this patient's own baseline about 219 minutes into monitoring. Upper-limb amplitude was at 29% of baseline at that point.
Similarity to this patient's own baseline fell as much as 92% while upper-limb amplitude rose to 1065% of baseline, finishing the case at about 716%. Arm strength was better at the first postop exam than before surgery. A fall in similarity together with a rise in amplitude can mean the signals are stronger than at baseline but shaped differently from it, which after decompression is consistent with improved conduction and strength rather than injury. This is a hypothesis from a small number of cases, not a validated indicator.
Similarity to this patient's own baseline fell as much as 49% while upper-limb amplitude rose to 586% of baseline, finishing the case at about 630%. Arm strength was better at the first postop exam than before surgery. A fall in similarity together with a rise in amplitude can mean the signals are stronger than at baseline but shaped differently from it, which after decompression is consistent with improved conduction and strength rather than injury. This is a hypothesis from a small number of cases, not a validated indicator.
Dates and identifying details have been removed. Pediatric cases were selected from the 14 patients with a true neurological change in a retrospective cohort, with times relative to the team alert. Adult cervical cases were selected from a retrospective cohort of 81 patients, with times measured from the first recorded signal set; the adult flag threshold was set on the same data. SureStim flag times come from a retrospective run of the model. Baseline similarity is measured on each patient's own early signals. These cases are not a measure of typical or prospective performance. SureStim is an investigational technology.
Partners and clinicians can request a full briefing.